It looks like some of our readers had the familiar error message with brown error 6084 06. There are a number of factors that can cause this issue. Let’s discuss it now.
Here is the solution to brown error 6084 21. It takes good mechanical skills, a small screwdriver, glue, and patience to fix this.
How to fix Braun blood pressure monitor 6084 error 21?
0 Here is the solution to error 21 brown 6084. To repair it you need a good clockwork, a small screwdriver, glue and patience. I have exactly the same error. I noticed that the pump motor does not start and error 21 is displayed on the screen! Opened again by unscrewing 2 screws.
I had the same flaw. I noticed that the pump motor is not running at full power and the screen shows error 21!
I opened it by unscrewing 2 screws. I clicked on the front and removed 2 more screws, possibly under the LCD screen. Unsolder the speaker wires. Disconnect 1 rubber hose that goes to thePressure sensor assembly and large white connector, and remove the circuit board. Now let’s look at the white pump motor. Unscrew 4 extremely black and individual ones. You can now see a white unbalanced wheel above the electric motor. This is problem! He’s stuck! Everything has gone too far here. You will definitely feel it when you try to move it (Brown’s design error). To reduce this … Remove the little white wheel, glue it again, but don’t press. Therefore, it can rotate freely. Make sure there is no excess epoxy in it and let it dry. You can apply 3V from an external power supply to test the motor pins and see if it works. Put it all together, but use your blood pressure. Good luck fixing this guys! 😉